Philips received a complaint on the heartstart xl defibrillator/monitor indicating that the ecg cables sheath is worn.Available details indicate that the device had exhibited symptoms of a failure.It was determined that this was a malfunction of the 989803145121,3 leadset, snap, iec, icu, which was replaced, and the device was returned to full functionality.The device remains at the customer site and no further evaluation is warranted at this time.Based on the information available and the testing conducted, the cause of the reported problem was a malfunction of the 989803145121,3 leadset, snap, iec, icu.The reported problem was confirmed.The customer replaced the 989803145121,3 leadset, snap, iec, icu to resolve the issue.It has been concluded that no further action is required at this time.
|