Model Number B35200
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
It was reported that a kink or bend was noted in extensions after doctor had difficulty pulling the excess into the pocket during the initial stage 2 procedure.Impedances checked with extensions connected and all were normal.Doctor decided to replace kinked/bent extensions and the issue was resolved.No symptoms were reported.
 
Manufacturer Narrative
Section d references the main component of the system.Other medical products in use during the event include: brand name sensight; product id b3400060m (serial: (b)(6); product type: 0191-extension; implant date (b)(6) 2023; explant date (b)(6) 2023, brand name sensight; product id b3400060 (serial: (b)(6); product type: 0191-extension; implant date (b)(6) 2023; explant date (b)(6) 2023.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3.Analysis of the extension (s/n (b)(6)) found no significant anomaly.Analysis of the extension (s/n (b)(6)) found no significant anomaly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPT
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18440139
MDR Text Key331876946
Report Number3004209178-2024-00153
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00763000642174
UDI-Public00763000642174
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberB35200
Device Catalogue NumberB35200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received01/03/2024
Supplement Dates Manufacturer Received01/26/2024
Supplement Dates FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10...
Patient Age64 YR
Patient SexMale
Patient Weight88 KG
-
-