• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number PLC161400
Device Problems Patient-Device Incompatibility (2682); Appropriate Term/Code Not Available (3191)
Event Date 12/24/2020
Event Type  Injury  
Event Description
On (b)(6) 2014, this patient underwent endovascular repair of an abdominal aortic aneurysm using gore excluder aaa endoprostheses.The patient tolerated the procedure.Pre-treatment cta showed that the maximum diameter of aortic aneurysm/lesion was 60.8mm.From (b)(6) 2014 to (b)(6) 2020 the follow up scans showed that the maximum diameter of the aortic aneurysm/lesion increased from 55mm to 107mm.Additionally, on (b)(6) 2020, a distal right limb type i endoleak was determined.It was reported that a loss of seal of the right distal iliac limb to the common iliac artery caused a distal type i endoleak.The patient had a severe tortuosity of the right iliac system.Reportedly, a distal type i endoleak caused an aneurysm enlargement.On (b)(6) 2021, the patient underwent endovascular procedure using gore excluder aaa endoprostheses (plc201400/ 2258291 and plc181400/ 22477819).The patient tolerated the procedure.
 
Manufacturer Narrative
H.6.Code c21: a review of the manufacturing records for the device is going to be conducted.The investigation is in process.The device remains implanted and is not available for analysis.Code a27 was used to cover the loss of seal of the right distal iliac limb to the common iliac artery.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
H6: code c19: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.The device remains implanted and is not available for analysis.Code a27 was used to cover the loss of seal of the right distal iliac limb to the common iliac artery.The patient had a severe tortuosity of the right iliac system.It was reported that a loss of seal of the right distal iliac limb to the common iliac artery caused a distal type i endoleak, and a distal type i endoleak caused an aneurysm enlargement.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention or additional intraoperative procedure time include but are not limited to endoleak and aneurysm enlargement.Per ifu, additional considerations for patient selection include but are not limited to: patient¿s anatomical suitability for endovascular repair.Ilio-femoral access vessel size and morphology (minimal thrombus, calcium and/or tortuosity) should be compatible with vascular access techniques.Additionally, the ifu states that an inadequate sealing may lead to endoleak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Type of Device
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
nataliya baramzina
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key18441447
MDR Text Key331863931
Report Number3007284313-2024-02988
Device Sequence Number1
Product Code MIH
UDI-Device Identifier00733132618729
UDI-Public00733132618729
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Catalogue NumberPLC161400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2023
Initial Date FDA Received01/03/2024
Supplement Dates Manufacturer Received01/08/2024
Supplement Dates FDA Received01/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Hospitalization; Required Intervention;
Patient Age58 YR
Patient SexMale
Patient Weight89 KG
Patient RaceWhite
-
-