Catalog Number 1829500
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  malfunction  
Event Description
It was reported that during a port placement procedure, the port attachment site allegedly could not be connected to the catheter.There was no reported patient injury.
 
Manufacturer Narrative
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a port placement procedure, the port attachment site allegedly could not be connected to the catheter.The procedure was complete by using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport duo mri implantable port was returned for sample evaluation.Functional, gross visual, tactile and microscopic visual evaluations were performed.Gap was noted between the two port stems, reflecting expansion.Both port septum was patent to infusion and aspiration without issue.No leaks or blockage was noted.No other anomalies were observed.Therefore, the investigation is confirmed for the reported loosening connection issue as gap was noted between the port stems.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 07/2024), g3, h2, h6 (method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H10: b5, d4 (expiry date: 07/2024), g3 h11: d4 (medical device lot number), h6 (method) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a port placement procedure, the port attachment site allegedly could not be connected to the catheter.The procedure was complete by using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18441674
MDR Text Key331986184
Report Number3006260740-2023-06008
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027185
UDI-Public(01)00801741027185
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K090512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1829500
Device Lot NumberREGZ3068
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2023
Initial Date FDA Received01/04/2024
Supplement Dates Manufacturer Received01/10/2024
03/29/2024
Supplement Dates FDA Received02/07/2024
04/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age25 YR
Patient SexMale
-
-