LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2023
Event Type  malfunction  
Event Description
Livanova deutschland has received a report of error/alarm shaft angel encoder fault.There is no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 double head pump the incident occurred in north kansas city, missouri.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The problem was traced back to a defective shaft encoder and was solved by part replacement.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
 
Manufacturer Narrative
The involved roller pump was manufactured in (b)(6) 2009 and according to the analysis of the complaints database no further events have been reported in the past.Based on the above facts and taking into account the investigation performed for similar events, the reported fault can be solely due to a defective shaft angle encoder.Taking into account the manufacturing year of the unit, it cannot be ruled out that wearing of the part that can be originated by the specific use at customer's facility may have contributed to the event.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18442171
MDR Text Key331873636
Report Number9611109-2024-00010
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received01/04/2024
Supplement Dates Manufacturer Received02/09/2024
Supplement Dates FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-