Model Number NX1000-16
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dyspnea (1816); Loss of consciousness (2418); Salivary Hypersecretion (4486)
Event Date 12/12/2023
Event Type  Injury  
Manufacturer Narrative
A review of the device history record (dhr) was conducted which confirmed that the device met all quality criteria and manufacturing specifications prior to release.There was no indication of a device malfunction from the available information.The nxstage system one user guide outlines risks associated with performing hemodialysis therapy and warns all treatments must be administered under a physician's' prescription and must be performed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.Udi: (b)(4).
 
Event Description
A report was received on 12 dec 2023 from the caregiver of a 61 year old male patient with a medical history including end stage renal disease, who stated the patient was dyspneic, foamed at the mouth and lost consciousness during a home hemodialysis treatment on (b)(6) 2023.Emergency medical services (ems) were called and the patient regained consciousness prior to being transported to the hospital.Additional information was received on 19 dec 2023 from the home therapy manager (htm), who stated the patient remained conscious during transfer and coded (nos) upon arriving at the hospital.The patient was admitted and evaluated for a stroke with no further details of hospitalization provided.Following hospital discharge, the patient has resumed therapy with the nxstage system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key18442436
MDR Text Key331865459
Report Number3003464075-2024-00001
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170469
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberNX1000-16
Device Catalogue NumberCYCLER VERSIHD 1.0, NO NIBP
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age61 YR
Patient SexMale
-
-