MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE VR; Pulse generator, permanent, implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE VR; Pulse generator, permanent, implantable Back to Search Results
Model Number CD1411-36Q
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
This report is to advise of an event observed during analysis.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The battery was analyzed by its manufacturer and an internal battery anomaly was found.The cause of premature battery depletion was due to the anomalous battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE VR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18443532
MDR Text Key331882649
Report Number2017865-2024-00225
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507738
UDI-Public05414734507738
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2019
Device Model NumberCD1411-36Q
Device Lot NumberS000041883
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/14/2023
Initial Date FDA Received01/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-