• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 00-1024-R
Device Problems Electrical /Electronic Property Problem (1198); Unable to Obtain Readings (1516)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Event Description
It was reported to philips that the etco2 is not working.No patient harm or injury has been reported.Further information will be send upon completion of the manufacturer's investigation.
 
Manufacturer Narrative
The reported complaint was confirmed as the co2 connector is physically damaged.The capnometer assembly was replaced to resolving the customer¿s complaint.Testing and verification was completed satisfactorily (calibrated nbp, co2 and touch screen) and the device was returned to service at the customer site.Based on the information available, no further action is necessary at this time.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS PRO
Type of Device
MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK  GU14 6XW
Manufacturer (Section G)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK   GU14 6XW
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key18445252
MDR Text Key331913117
Report Number3003832357-2024-00008
Device Sequence Number1
Product Code MHX
UDI-Device Identifier05060472441027
UDI-Public05060472441027
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201746
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Remedial Action Repair
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-1024-R
Device Catalogue Number989706000051
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received01/04/2024
Supplement Dates Manufacturer Received12/21/2023
Supplement Dates FDA Received02/07/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-