Model Number 176630
Device Problems Mechanical Problem (1384); Misfire (2532); Mechanics Altered (2984)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic procedure, the device misfired.
 
Manufacturer Narrative
Additional information: a3, a4, b5, g3, h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic inguinal hernia repair procedure, when the device was positioned to be fired, the clip did not form, and just fell out of the end of the device.The surgeon attempted to use it again, the device was fired but no clips appeared on the end of the device.The surgeon double tapped and a clip fell out.The next firing appeared normal however when the device was positioned and fired the clip ¿shot¿ off the end like a projectile.Another device was used to resolve the issue.There was no patient injury.
 
Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.It was reported that there was a reported condition of misfire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if force was applied while rotating the shaft during application and also if the distal end of the device was subjected to excessive manipulation during application of the instrument.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: firing a clip over another clip or other obstructions may result in bleeding and/or leakage and may damage the instrument jaws.Also, avoid excessive twisting of the jaws or tissue manipulation when firing the instrument.Deflecting the jaws and/or shaft during firing may result in an improperly formed clip and possible bleeding and/or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18446364
MDR Text Key331937847
Report Number9612501-2024-00044
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057852
UDI-Public10884521057852
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K100242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number176630
Device Catalogue Number176630
Device Lot NumberJ3J3839Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/11/2023
Initial Date FDA Received01/04/2024
Supplement Dates Manufacturer Received12/21/2023
02/15/2024
Supplement Dates FDA Received01/12/2024
02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
Patient Weight52 KG
-
-