Model Number PM2152
Device Problems High impedance (1291); Defective Device (2588)
Patient Problem Hematoma (1884)
Event Date 12/13/2023
Event Type  Injury  
Event Description
It was reported that an increase in the pacing impedance was observed on the device.Additionally, a hematoma was formed on the patient.As a result, the device was explanted and replaced to resolve the event.The cause of the event was alleged to be due to a suspected header anomaly.The patient was explanted and replaced to resolve the event.
 
Manufacturer Narrative
The reported event of high atrial impedance was confirmed.Analysis revealed the atrial contact spring was missing from the ring slot of the atrial connector port which resulted in the high atrial impedance.The connector port dimensions were normal.No device anomalies were revealed except for the missing spring.The missing spring was due to a manufacturing anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY¿ CORE DR, JPD
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18446725
MDR Text Key332144815
Report Number2017865-2024-00374
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM2152
Device Lot NumberP000184478
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/28/2023
Initial Date FDA Received01/04/2024
Supplement Dates Manufacturer Received09/23/2015
Supplement Dates FDA Received05/31/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-