Catalog Number 2227
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Additional information: h6 component code: g07002 - device not returned.Additional information provided: the hole was found before using the product and the product was not used on the patient.The blood found on the complaint product was from the blood on the surgeon's glove, which also adhered to the product.Additional information has been requested and received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.Was the drain activated?unk.If yes, was leakage detected?not reported.Was another drain needed to correct the situation?unk.If yes, was the new drain placed surgically during a second procedure? the device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported a patient underwent an unknown procedure on an unknown date and a drain was used.There was a hole in the drain outside the body there were no adverse consequences to the patient.Additional information has been requested.
 
Manufacturer Narrative
Product complaint # (b)(4).Additional information: h6 type of investigation.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).H3 evaluation: complaint sample review : one complaint sample of drain with trocar was received for evaluation, no negative observation was identified while product was inspected visually.Also, no hole in the drain was found against complaint statement 'hole in the drain outside the body'.Documents review : during investigation of complaint, batch review of in-process and final packaging inspections, as well as the manufacturing process is performed, these products are reported to be manufactured, inspected, and packaged in conformance with customer's specifications and have been found to be acceptable within all parameters.No discrepancy as per the reported defect is observed.Retention sample review : no negative observation was found.As per standard practice, 100% functional test and 100% visual inspection was carried out visually before and after packing of finished goods, prior to the product release.There was no scope to miss such defect, at manufacturing / release stage.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLAKE SI DRAIN HBLS 10FR RND W/ 1/8IN
Type of Device
CATHETER, IRRIGATION
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
DEGANIA INDIA
251, sector-6, imt manesar
gurugram 12205 2
IN   122052
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key18448866
MDR Text Key332160309
Report Number2210968-2024-00109
Device Sequence Number1
Product Code GBX
UDI-Device Identifier10705031003552
UDI-Public10705031003552
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
CL I EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2227
Device Lot NumberJ2300036
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/08/2023
Initial Date FDA Received01/04/2024
Supplement Dates Manufacturer Received12/13/2023
01/20/2024
Supplement Dates FDA Received01/11/2024
02/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-