Model Number U125
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Patient Problems Dyspnea (1816); Asystole (4442)
Event Date 12/12/2023
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) recorded episodes with noise over sensing at both right atrial and right ventricular channels.After analyzing the noise, it appears as electromagnetic interference more than minute ventilation termination algorithm.Also, no signal artifact monitoring episode was recorded and no out of range lead impedances were recorded.It was suspected that a procedure could have taken place.The right atrial rhythm appears to be marching through, but it was not possible to determine if the right ventricle showed appropriate pacing.The patient was hospitalized at the moment of the call due to shortness of breath.The crt-p remains in service at this time.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18451340
MDR Text Key332178285
Report Number2124215-2024-00371
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/23/2023
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number721436
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2023
Initial Date FDA Received01/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age88 YR
Patient SexFemale
-
-