Catalog Number 2000-9061 |
Device Problem
Fracture (1260)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 12/07/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Phone number:(b)(6).Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
|
Event Description
|
It was reported that the tip of a polaris plug driver fractured during surgery.There was no reported patient harm; however, the patient retained the tip.
|
|
Event Description
|
It was reported that the tip of a polaris plug driver fractured during surgery.There was no reported patient harm; however, the patient retained the tip.
|
|
Manufacturer Narrative
|
Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip has fractured off.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: per dhr review, the part was likely conforming when it left highridge control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
|