Catalog Number 2000-9061
Device Problem Fracture (1260)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/07/2023
Event Type  Injury  
Manufacturer Narrative
Phone number:(b)(6).Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that the tip of a polaris plug driver fractured during surgery.There was no reported patient harm; however, the patient retained the tip.
 
Event Description
It was reported that the tip of a polaris plug driver fractured during surgery.There was no reported patient harm; however, the patient retained the tip.
 
Manufacturer Narrative
Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip has fractured off.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: per dhr review, the part was likely conforming when it left highridge control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUG DRIVER
Type of Device
SCREWDRIVER
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
sabrina abla
10225 westmoor dr.
westminster, CO 80021
7206965158
MDR Report Key18453838
MDR Text Key332257001
Report Number3012447612-2024-00404
Device Sequence Number1
Product Code HXX
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
EXEMPT
Exemption Number5645646
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2000-9061
Device Lot NumberZB170803
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age56 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-