ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2023
Event Type  malfunction  
Event Description
A non-healthcare professional reported during the intraocular lens (iol) implant procedure the iol was scratched following the implantation.There was no patient harm.Additional information has been requested.There are three medical device reports associated with this file.This report is associated with the 1 of 3 files.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Used company cartridge was returned.Inadequate viscoelastic was observed in the device.The tip had moderate stress.The cartridge had evidence of placement into a handpiece.The used company cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.A qualified lens model/diopter was indicated with a non-qualified viscoelastic.The handpiece used was not provided.It is unknown if a qualified handpiece was used.No problem was found with the returned company cartridge.Top coat dye stain testing was conducted with acceptable results.The lens damage could not be verified.The lens was not returned for evaluation.The root cause for the reported ¿lens scratched¿ may be related to a failure to follow the ifu (instructions for use).Inadequate viscoelastic was observed in the returned used company cartridge.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.In addition, a non-qualified viscoelastic was indicated.It is unknown if a qualified handpiece was used.The ifu instructs: the company cartridges are qualified for use with compatible company handpieces for the surgical implantation of company qualified foldable iols.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18453867
MDR Text Key332260052
Report Number1119421-2024-00033
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot Number15682031
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 12/12/2023
Initial Date FDA Received01/05/2024
Supplement Dates Manufacturer Received01/31/2024
02/28/2024
Supplement Dates FDA Received02/28/2024
03/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL; EYEFILL C 207005
Patient Age62 YR
Patient SexFemale
-
-