Model Number N/A |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/10/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during cleaning that the coupler failed and needed replacement.Upon evaluation of the device, it was determined that there was a small puff of smoke coming from the coupler connection on the board.There was no harm, injury or delay as there was no patient involvement.Due diligence is complete.No additional information is available.No adverse event reported.
|
|
Manufacturer Narrative
|
This event is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined evac station working intermittently.Noticed a small puff of smoke coming from the coupler connection on the board.Coupler was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|