Model Number BB1001
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418)
Event Date 12/19/2023
Event Type  Injury  
Manufacturer Narrative
No product was returned for evaluation.The ilet logs were reviewed by beta bionics failure investigation department.No product performance issues were identified.If the product is received at a later date, the complaint will be reopened and investigated accordingly.No anomalies were observed.The device history record (dhr) review was completed, and this device passed all manufacturing release criteria for distribution.There were no issues identified that would have impacted this event.
 
Event Description
On 12/20/23 an ilet patient reported he was driving on 12/19/2023 at around 4:20-4:30pm.The patient noticed his blood glucose (bg) levels drop from 136mg/dl down to 32mg/dl rapidly.The patient stated he lost consciousness while driving and woke up in the hospital.During his hospital visit, he was administered an iv drip of glucagon.The patient was able to get discharged that evening of (b)(6) 2023.The patient is currently not using the ilet and will return using his previous insulin pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILET BIONIC PANCREAS
Type of Device
INTEROPERABLE AUTOMATED GLYCEMIC CONTROLLER
Manufacturer (Section D)
BETA BIONICS
11 hughes
irvine CA 92618
Manufacturer (Section G)
BETA BIONICS
11 hughes
irvine CA 92618
Manufacturer Contact
joe conkey
11 hughes
irvine, CA 92618
MDR Report Key18456625
MDR Text Key332211314
Report Number3019004087-2024-00005
Device Sequence Number1
Product Code QJI
Combination Product (y/n)N
PMA/PMN Number
K232224
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberBB1001
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received01/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient SexMale
-
-