PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD20
Device Problem Radiation Output Failure (4027)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2023
Event Type  malfunction  
Event Description
It was reported to philips that the system will not x-ray.The device was in clinical use.No patient or user harm was reported.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the procedure got delayed.The philips field service engineer (fse) inspected the system onsite and confirmed that the system wouldn't make an x-ray.Upon troubleshooting, fse found that the issue was due to defective ippc.To resolve the issue, fse replaced ippc, hard drives, loaded software and returned ippc to philips for further analysis.The analysis confirmed the malfunction of the ippc due to failed pci (peripheral connect interface).Following the replacement of ippc, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
derrick massuri
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18463063
MDR Text Key332275327
Report Number3003768277-2024-00082
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059054
UDI-Public00884838059054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/09/2023
Initial Date FDA Received01/08/2024
Supplement Dates Manufacturer Received02/22/2024
Supplement Dates FDA Received02/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse