The device was not returned to zoll medical corporation for evaluation.Instead, the data file of the customer's report was provided.Review of the data file indicated cpr was being performed during the analysis and contributed to the shock advised result.The manual states to "keep patient motionless during ecg analysis.Do not touch the patient during analysis.Cease all movement via stretcher or vehicle before analyzing the ecg".Based on our review of the data, we have concluded that the analysis program results were correct for the specific analyses in question.No trend is associated with reports of this type.
|