ZOLL MEDICAL CORPORATION AED PLUS; DEFIBRILLATOR Back to Search Results
Model Number AED PLUS
Device Problem Incorrect Interpretation of Signal (1543)
Patient Problem Insufficient Information (4580)
Event Date 11/23/2023
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a 69-year-old patient (gender unknown), the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable.Complainant indicated that the patient subsequently expired.
 
Manufacturer Narrative
The device was not returned to zoll medical corporation for evaluation.Instead, the data file of the customer's report was provided.Review of the data file indicated cpr was being performed during the analysis and contributed to the shock advised result.The manual states to "keep patient motionless during ecg analysis.Do not touch the patient during analysis.Cease all movement via stretcher or vehicle before analyzing the ecg".Based on our review of the data, we have concluded that the analysis program results were correct for the specific analyses in question.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED PLUS
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18463180
MDR Text Key332266358
Report Number1220908-2023-05104
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011541/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberAED PLUS
Device Catalogue NumberAED PLUS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/18/2023
Initial Date FDA Received01/08/2024
Supplement Dates Manufacturer Received12/18/2023
Supplement Dates FDA Received03/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age69 YR
-
-