• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II XENON LIGHT SOURCE Back to Search Results
Model Number CLV-180
Device Problem Flare or Flash (2942)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, the light source front panel leds (light-emitting diodes) were blinking continuously.The issue was found during preventive maintenance and no patient was involved.There was no reports of patient harm or impact associated with this event.
 
Manufacturer Narrative
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if additional information is provided by the user facility.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation and the device evaluation.Information added to the fields: d8, d9, h3, h4, and h6.The device was evaluated by olympus.The evaluation found that the allegation that the leds were blinking continuously was confirmed.Additional issues were identified during the device evaluation: the two towers with the mechanism were locked; the inside motor and reducer gears were detected to have lost grease effectiveness and have hardened.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the root cause of the event could not be determined.However, it is likely that front panel illumination blinking occurred because the grease on the filter turret motor lost its effect and malfunctioned due to hardening.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18463879
MDR Text Key332974730
Report Number3002808148-2024-00223
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170216961
UDI-Public04953170216961
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/17/2023
Initial Date FDA Received01/08/2024
Supplement Dates Manufacturer Received01/24/2024
Supplement Dates FDA Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-