Model Number U225
Device Problems Signal Artifact/Noise (1036); Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439); Off-Label Use (1494); Under-Sensing (1661); High Capture Threshold (3266)
Patient Problem Asystole (4442)
Event Date 12/18/2023
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) system had shown right ventricular (rv) lead noise in the past, and the rv lead and the left ventricular (lv) lead ports were switched.During recent generator change, device measurements were tested, and it was identified that when not switched, cross chamber blanking, and pacing inhibition in the original lv lead were noted.During system evaluation, the original rv lead, now in the lv port was exhibiting loss of capture at high capture thresholds.In addition, when using the correct intended port, asystole was experienced.Reprogramming options were discussed.This crt-p system remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST CRT-P
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18465588
MDR Text Key332397148
Report Number2124215-2024-00595
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559426
UDI-Public00802526559426
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberU225
Device Catalogue NumberU225
Device Lot Number727507
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age98 YR
Patient SexFemale
-
-