• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II CR LOCK FEM IMPLANT IMPACTOR; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II CR LOCK FEM IMPLANT IMPACTOR; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 74011711
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2023
Event Type  malfunction  
Event Description
It was reported that, during a tka surgery, a jrny ii cr lock fem implant impactor broke during a case.It is unknown if the breakage occurred on the impactor itself, or on the bumper.It is unclear how this surgery was completed.No delay or injuries were reported.
 
Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Manufacturer Narrative
Section h3, h6: the devices were not returned for evaluation; therefore, a device analysis could not be performed.The devices' batch numbers were not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed devices, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.At this time, we have no evidence to conclude that the products failed to meet any specifications at the time of manufacture.These devices are reusable instruments that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II CR LOCK FEM IMPLANT IMPACTOR
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18465850
MDR Text Key332749674
Report Number1020279-2024-00047
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556422892
UDI-Public00885556422892
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121443
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number74011711
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/15/2023
Initial Date FDA Received01/08/2024
Supplement Dates Manufacturer Received01/29/2024
Supplement Dates FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
JRNY II CR FEM IMP BUMPER RT(PN:74011857)
-
-