Model Number PM2240
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital for a scheduled implant procedure.During the procedure, there was difficulty to remove the right ventricular (rv) lead from the device.The pacemaker was then successfully removed and replaced.No patient condition was reported.
 
Manufacturer Narrative
The reported event of difficulty to remove the ventricle lead from the header was confirmed.Final analysis revealed there was blood accumulation in the connector port zones including the connector block areas.The blood in all these areas had a bonding affect between the inside of the connector ports and the silicone lead body surfaces of the lead and also the lead pin.This made the removal of the lead difficult.No device anomalies were found.The blood was most likely not cleaned from the proximal end of the lead before it was inserted into the connector at time of implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18472373
MDR Text Key332509376
Report Number2017865-2024-00724
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507073
UDI-Public05414734507073
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2016
Device Model NumberPM2240
Device Lot NumberA000001270
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received01/08/2024
Supplement Dates Manufacturer Received01/17/2024
Supplement Dates FDA Received01/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age96 YR
Patient SexFemale
Patient Weight37 KG
-
-