MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR TRUWAVE DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR TRUWAVE DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number PX260
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2023
Event Type  malfunction  
Manufacturer Narrative
Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
As reported, during use in patient with this disposable pressure transducer, the mean arterial pressure (map) was inaccurately provided.No error message was displayed and the patient was not treated according to the inaccurate values.Patient demographics unable to be obtained.There was no allegation of patient injury.The device was available for evaluation.
 
Manufacturer Narrative
The disposable pressure transducer (dpt) was received for a full examination.The report of inaccurate values was not able to be confirmed.Dpt zeroed and sensed pressure accurately on pressure monitor.Pressure did not show any drift during output drift testing and met specification.Electrical testing showed that both input impedance and output impedance were within specifications.Zero-offset also met specification.No leakage or occlusion was detected from the kit during pressure test.No visible damage was observed from the kit.The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.Based on further engineering investigation, it was verified that there are manufacturing controls to avoid potential causes related to the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUWAVE DISPOSABLE PRESSURE TRANSDUCER
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key18474188
MDR Text Key332424215
Report Number2015691-2024-00176
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K222216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPX260
Device Lot Number64808039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/17/2023
Initial Date FDA Received01/09/2024
Supplement Dates Manufacturer Received03/11/2024
Supplement Dates FDA Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-