BAXTER HEALTHCARE CORPORATION NON VENTED HIGH VOL.INLET,N/S; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938173
Device Problems Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2023
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a non-vented high-volume inlet separated from the bag and caused product to spill on the pec surface.This occurred prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Correction: d1, d4.H11: the actual device was not available; however, a photograph of the sample was provided for evaluation.The returned photograph was reviewed, and it was noted that there was liquid in the tubing and the tubing was separated from the spike.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NON VENTED HIGH VOL.INLET,N/S
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
PLASTICOS Y MATERIAS PRIMAS PYMPSA
3609-2 juan de la barrera
parque industrial el alamo
guadalajara, jalisco 44490
MX   44490
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18476322
MDR Text Key332894377
Report Number1416980-2023-07008
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412475783
UDI-Public(01)00085412475783
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K002705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938173
Device Lot Number803579
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/18/2023
Initial Date FDA Received01/09/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-