• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS INFINITY LAB CORE LICENSE 3.X; CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS INFINITY LAB CORE LICENSE 3.X; CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE Back to Search Results
Catalog Number 09180443001
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2023
Event Type  malfunction  
Event Description
We received an allegation of a software malfunction in the cobas infinity lab core license 3.X.The reporter stated that the instrument driver was saving a result in the field "r4" instead of the field "r4-1"; the driver was saving "62.42^" instead of "62.42".There was no report of any patient harm because of the event.
 
Manufacturer Narrative
The software version is 3.03.12.The investigation is ongoing.
 
Manufacturer Narrative
The investigation determined that the event occurred because the workflow manager driver was not able to capture the information as defined in the host interface manual of the instrument.The investigation verified that the event was a software issue in the cobas infinity.This issue has been addressed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS INFINITY LAB CORE LICENSE 3.X
Type of Device
CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18476660
MDR Text Key332998065
Report Number1823260-2024-00099
Device Sequence Number1
Product Code JQP
UDI-Device Identifier07613336177129
UDI-Public07613336177129
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09180443001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/15/2023
Initial Date FDA Received01/09/2024
Supplement Dates Manufacturer Received01/18/2024
Supplement Dates FDA Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-