|
Model Number UNVCA12STF |
Device Problems
Leak/Splash (1354); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
D10 concomitant products: unvca12stf, unvca12stf universal 12 stdfix cannula, (lot #j3h2636y) unvca12stf, unvca12stf universal 12 stdfix cannula, (lot #j3h2636y) unvca12stf, unvca12stf universal 12 stdfix cannula, (lot #j3h2636y) unvca12stf, unvca12stf universal 12 stdfix cannula, (lot #j3h2636y) unvca12stf, unvca12stf universal 12 stdfix cannula, (lot #j3h2636y); medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic fundoplication, all six devices had leakage.The surgical time was extended by 30 minutes or more due to the product problem as they had give more co2.The device was changed to resolve the issue.
|
|
Manufacturer Narrative
|
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the trocar and cannula appeared intact.The circular seal and duckbill seal appeared intact.Functionally, the device passed an air leak test.The duckbill seal failed during manual manipulation using an endo peanut.The trocar was broken open to remove the circular seal to check for subassembly overall height.It was reported that the device had a leakage.The reported issue was confirmed.The most likely cause was traced to component failure.It was also reported that medical complication was reported.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|