Trackwise#: (b)(4) updated sections: b4, g4, g7, h2, h3, h6, h10 the device was returned to the factory for evaluation on (b)(6) 2024.An investigation was conducted on (b)(6) 2024.A visual inspection was conducted.Both the harvesting device as well as the cannula was returned for evaluation.Signs of clinical use and evidence of blood was observed on the intact cannula and c-ring.The gray silicone insulation on the tip of the cold jaw was observed to be peeled back, exposing the cold metal tip of the jaw.The heater wire was observed to be intact with no visual defects observed.An electrical evaluation was conducted.A pre-cautery test was performed per the direction for use (dfu) with the reference cable, adapter and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.An activation and transection capability test was performed over four (4) repetitions using "max life test method (b)(4) rev aa.The device successfully transected tissue four (4) times.Based on the returned condition of the device, and the no specific failure reported, the analyzed failure "peeled; delaminated; jaw" was observed.The lot # 3000305725 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|