• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA DR ICD, US; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA DR ICD, US; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number CD2357-40C
Device Problem Failure to Convert Rhythm (1540)
Patient Problem Discomfort (2330)
Event Date 12/19/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for follow up after the implantable cardioverter delivered a shock for an episode of ventricular fibrillation.The first shock failed to convert the patient back to sinus rhythm.However the second shock at maximum energy was successful.The second shock caused the patient pain, and it was discussed that the patient had high defibrillation threshold.No interventions were made.The patient was stable.
 
Manufacturer Narrative
Additional information: a4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18478401
MDR Text Key332725687
Report Number2017865-2024-00792
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508063
UDI-Public05414734508063
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2018
Device Model NumberCD2357-40C
Device Lot NumberA000033575
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received01/09/2024
Supplement Dates Manufacturer Received01/09/2024
Supplement Dates FDA Received01/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RIGHT VENTRICULAR LEAD; TENDRIL
Patient SexMale
Patient Weight104 KG
-
-