ABBOTT MEDICAL REGENT HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number UNK REGENT HEART VALVE
Device Problems Fracture (1260); Detachment of Device or Device Component (2907)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/15/2023
Event Type  Injury  
Event Description
It was reported that on (b)(6) 2023, an unknown regent valve was chosen for implantation.After implant, when the valve was rotated it broke, resulting in one of the leaflets falling into the ventricle.The leaflet was retrieved and a replacement mechanical heart valve was implanted.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
An event of leaflet detachment and fracture was reported.The possible causes of the reported event include an external force applied to the valve leaflets that may cause structural damage, the use of a hard or rigid instruments to test leaflet mobility and oversizing of the valve.Field indicated that the valve was rotated during implantation which could have contributed to the reported event.Based on available information and without the device to analyze, the root cause of the reported event could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18480273
MDR Text Key332491240
Report Number2135147-2024-00162
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK REGENT HEART VALVE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/09/2024
Supplement Dates Manufacturer Received03/05/2024
Supplement Dates FDA Received03/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-