• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, DR PRESSURE MONITORING SET; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, DR PRESSURE MONITORING SET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number DPTUNKNOWN
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The 32 of 35 devices are not available for return.Without the return of the product, it is not possible to determine if damages or defects exist on the product, nor can any manufacturing nonconformance, failure mode, root cause, or potential contributing factors be identified.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.Complaint histories for all reported events are reviewed against trending control limits, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that icu staff have had discrepancies between arterial line readings and manual cuff readings since 05dec2022.The values on the arterial line were consistently higher compared to the manual cuff readings.There were at least 35 patients with invasive blood pressure issues -ibp, and each were using a truwave transducer.Different lots were used, and numbers are unknown, but models used are px284 and pxvp2284, quantity of each is unknown, patient demographics and date of occurrences are not provided.Staff tried to troubleshoot by ensuring there was no air in tubing or transducer, the stopcocks were in the appropriate positions, the connections were tight without cracks or leaks, the pressure bag has normal saline and was inflated to 300mmhg, the transducer was leveled and re-zeroed, the line was aspirated and flushed, excess tubing was removed, and that the square wave was checked.The kits were primed via gravity.There was no abnormal wave form or error messages.Hospital investigated the issue themselves.In the investigation, troubleshooting was provided to swap the ibp cable and nibp tubing with the cuff and the issue was not resolved.The truwave transducers were then tested in the biomed shop and showed accurate pressures.Philips tested icu rooms with biomed configuration.All tests came back with no issues and confirmed the settings were good.Nibp was tested with biomed and passed all functional tests.Manual bp checks were also performed by nurses and matched the philips monitor.A zoll defibrillator, hemosphere, and philips transport monitor were next tested.Ibp and nibp matched one of the icu patient monitors.Between the different monitors, the transducer was the common denominator.All events occurred during use.There were no patient injuries, but unintended treatments such as prn anti-hypertensive medications were administered.Of the 35 devices being reported, only 3 will be returned and it is unclear which models will be returned.This report is for the 32 devices not being returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING SET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES, DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES, DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR  
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
9492507258
MDR Report Key18480912
MDR Text Key333009670
Report Number2015691-2024-00207
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDPTUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2023
Initial Date FDA Received01/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-