MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT Back to Search Results
Model Number MR850
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A healthcare facility in new york reported an issue with an mr850 respiratory humidifier.Upon device servicing at the regional office in california, it was identified that the power cord was damaged and the copper wires were visible and exposed.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr850 respiratory humidifier was returned to our fisher & paykel (f&p) healthcare service centre in california where it was inspected by a trained f&p service technician.Results: visual inspection of the returned power cord identified that the power cord insulation was cut and the copper wires were visible and exposed.Conclusion: we are unable to determine the cause of the observed damage.However, it is likely due to physical damage during use.During production the electrical connections of the earth wires on all mr850 units are 100% tested for electrical continuity.Any product that fails is rejected.All mr850 units are visually inspected for damaged power cords before release for distribution.This suggests that the damage occurred after it had been distributed.The mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking, performance and electrical safety testing of the mr850 heater base.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 respiratory humidifier is complaint with the following electrical standards: as/nzs 3200.1.0, can/csa 22.2 no.601.01, ul 60601-1, iec 60601-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRATORY HUMIDIFIER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key18480925
MDR Text Key332669848
Report Number9611451-2024-00010
Device Sequence Number1
Product Code BTT
UDI-Device Identifier09420012407290
UDI-Public(01)09420012407290(10)2101501594(11)210209
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR850
Device Catalogue NumberMR850
Device Lot Number2101501594
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2023
Initial Date FDA Received01/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/09/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-