Model Number 3100A |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire medical file identification: (b)(4).H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not returned yet.
|
|
Event Description
|
It was reported to vyaire medical that the 3100a ventilator has a leak in the gas supply by hissing and making a very loud noise from the housing.Metal pipes get very cold.As of this time there is no information about patient involvement associated with this reported event.
|
|
Manufacturer Narrative
|
Result of investigation: the device was opened and fully inspected.No leaks were found.Subsequently, an stk (safety test) was carried out.The reported complaint was not reproducible.Additional information: the customer indicated that there was no patient involvement with the incident.
|
|