OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 6,4/7,8 FR. X 430 MM, 7°, STRAIGHT OCULAR ANGLE, 4,2 FR. CHANNEL
|
Back to Search Results |
|
Model Number WA29048A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1: (b)(6) healthcare corporation.The device was returned to olympus for evaluation and the evaluation found moisture under the distal cover glass, a bent outer tube, dents on the distal end, moisture under the cover glass on the objective window, a missing color ring on the eyepiece unit, and a shadow on the image due to a bent outer tube.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus, the eyepiece came off the ureteroscope.The issue was found during reprocessing.There were no reports of patient harm.
|
|