SENSEONICS, INCORPORATED EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 101967-710
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 04/14/2020
Event Type  Injury  
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.Bleeding at insertion site is a known and anticipated potential adverse effect associated with sensor insertion.User was taken to the emergency room where the incision site was closed with 2-3 stitches and tegaderm.The heavy bleeding was likely due to the blood thinner the user had been taking since (b)(6) 2020.The suture resolved the bleeding, and the user was fine afterwards.
 
Event Description
Senseonics was made aware of an incident where the patient experienced excessive bleeding at the insertion site soon after insertion.The patient's wife found that the patient's shirt and trousers were full of blood, so she applied a pressure bandage over the insertion site.The bleeding continued, so the patient went to the emergency room.A surgeon at the er recommended removing the sensor, which the patient did not want, so the insertion site was sutured instead with tegaderm placed on top.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS, INCORPORATED
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS, INCORPORATED
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key18488912
MDR Text Key332588352
Report Number3009862700-2023-00492
Device Sequence Number1
Product Code QCD
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date07/10/2020
Device Model Number101967-710
Device Catalogue NumberFG-4400-10-302
Device Lot NumberWP06494
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/10/2020
Initial Date FDA Received01/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-