ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0705-XTW
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
It was reported that a patient presented with grade 4 mixed mitral regurgitation (mr) for a mitraclip procedure.During use of the first clip, the anterior gripper could not be lowered.The gripper lowered for the first capture attempt, and was raised and could not lower with the second attempt.The clip was locked and standard troubleshooting was performed.The gripper could not lower, so the clip was removed and replaced with new one.During post-procedure handling, it was noticed that the gripper would not lower from the shaft.The physician manually pulled on the stuck gripper and it lowered.Once freed manually it worked intermittently but not normally (continued to get stuck).All steps were followed per instructions for use (ifu) and no other incidents occurred.The procedure was completed successfully with a mr grade of 1.There were no adverse patient effects or clinically significant delay.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and without the device to analyze, a cause of the reported difficult to open or close (gripper actuation - single) could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18488963
MDR Text Key332653569
Report Number2135147-2024-00185
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0705-XTW
Device Lot Number30731R1040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received01/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
-
-