Catalog Number J-NCS-504070
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.G4: pma/510(k)# = k180552 this report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
As reported, prior to use, a hair was found inside the packaging of a modified novy cornual cannulation set.Another device was used to complete the procedure.The patient did not require any additional procedures or experience any adverse effects due to this occurrence.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary as reported, prior to use, a hair was found inside the packaging of a modified novy cornual cannulation set.Another device was used to complete the procedure.The patient did not require any additional procedures or experience any adverse effects due to this occurrence.Investigation ¿ evaluation a document-based investigation was performed including a review of complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions (mi), and quality control procedures, as well as a visual inspection and functional test of the device were conducted.One modified novy cornual cannulation set was returned to cook for evaluation in partially opened packaging.A hair in was observed in the package.A document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot revealed no recorded non-conformances relevant to the failure mode.Additionally, there have been no other complaints filed for this lot.Therefore, there is no evidence to suggest that nonconforming product are in the field.The ifu supplied with the device, j-ncs-504070, was reviewed and includes the following how supplied do not use the product if there is doubt as to whether the product is sterile.Based on the information provided, inspection of the returned device, and the results of the investigation, cook has concluded the cause of the complaint is manufacturing.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODIFIED NOVY CORNUAL CANNULATION SET
Type of Device
MOV CATHETER, SALPINGOGRAPHY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC.
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18489295
MDR Text Key333035171
Report Number1820334-2024-00057
Device Sequence Number1
Product Code MOV
UDI-Device Identifier00827002175587
UDI-Public(01)00827002175587(17)250323(10)14617737
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberJ-NCS-504070
Device Lot Number14617737
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/19/2023
Initial Date FDA Received01/10/2024
Supplement Dates Manufacturer Received01/24/2024
Supplement Dates FDA Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age34 YR
Patient SexFemale
Patient Weight45 KG
-
-