• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number LDA210Q/58
Device Problems Difficult to Fold, Unfold or Collapse (1254); Retraction Problem (1536); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  Injury  
Event Description
It was reported, that the patient presented in clinic for a follow up.Device interrogation revealed, high capture thresholds.And difficult extending and retracting the helix on the right ventricular (rv) lead.The physician elected to explant and replace the rv lead.The patient was in stable condition.
 
Manufacturer Narrative
The reported events were high capture threshold and helix mechanism issue.As received, a complete lead was returned in one piece for analysis.As received, a complete lead was returned in one piece.The cause of the reported helix event was isolated to an over torqued inner coil, consistent with procedural damage and a clogged helix.No other anomalies were seen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18490016
MDR Text Key332601325
Report Number2017865-2024-00909
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734507325
UDI-Public05414734507325
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLDA210Q/58
Device Lot NumberA000141043
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received01/10/2024
Supplement Dates Manufacturer Received02/07/2024
Supplement Dates FDA Received02/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CDVRA500Q
Patient Outcome(s) Required Intervention;
Patient Age58 YR
Patient SexMale
Patient Weight82 KG
-
-