ST. JUDE MEDICAL, INC.(CRM-SYLMAR) OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES
|
Back to Search Results |
|
Model Number LDA210Q/58 |
Device Problems
Difficult to Fold, Unfold or Collapse (1254); Retraction Problem (1536); Capturing Problem (2891)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/21/2023 |
Event Type
Injury
|
Event Description
|
It was reported, that the patient presented in clinic for a follow up.Device interrogation revealed, high capture thresholds.And difficult extending and retracting the helix on the right ventricular (rv) lead.The physician elected to explant and replace the rv lead.The patient was in stable condition.
|
|
Manufacturer Narrative
|
The reported events were high capture threshold and helix mechanism issue.As received, a complete lead was returned in one piece for analysis.As received, a complete lead was returned in one piece.The cause of the reported helix event was isolated to an over torqued inner coil, consistent with procedural damage and a clogged helix.No other anomalies were seen.
|
|