BIOSENSE WEBSTER INC CARTO® 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Catalog Number FG540000
Device Problems Overheating of Device (1437); Fitting Problem (2183); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2023
Event Type  malfunction  
Manufacturer Narrative
The hardware investigation has begun but it has not been completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a carto® 3 system and they noticed black char on the inside of the connector and the inside of the port.It was reported that there were erratic temperature readings on the ngen generator when the catheter was connected to the patient interface unit (piu).Caller disconnected the catheter cable from the piu and they noticed some water in the connector of the cable and the port on the piu.The cable was replaced and there were no temperature readings on the ngen generator.When the cable was disconnected from the piu, they noticed black char on the inside of the connector and the inside of the port.Caller swapped out the piu and the location pad and continued the procedure.Caller confirmed there was some "moisture" visible in 20 pole b port on the piu.Caller stated it appeared when flushing the smartablate tubing, the port on the tubing was facing the piu and it may have caused some water damage.The ablation catheter was never used for ablation.Erratic temperatures were noticed while it was outside the body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO® 3 SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER (ISRAEL) LTD.
4 hatnufah street
yokneam 20667 17
IS   2066717
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18490387
MDR Text Key333014561
Report Number2029046-2024-00145
Device Sequence Number1
Product Code DQK
UDI-Device Identifier10846835000870
UDI-Public10846835000870
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133916
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFG540000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/19/2023
Initial Date FDA Received01/10/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/29/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
NGEN RF GENERATOR, US.; THMCL SMTCH SF BID, TC, F-J.; UNK_SMARTABLATE PUMP TUBING.
-
-