Catalog Number 112844110 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Pain (1994); Loss of Range of Motion (2032); Unspecified Tissue Injury (4559)
|
Event Date 01/02/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that the patient had her shoulder replaced approximately in 2010 by surgeon where she had a global advantage shoulder prosthesis.The patient did well until the last year where she has cuff tear arthropathy and lost mobility and became painful.Surgeon and the patient made the decision to explant the global construct and implant an inhance reverse construct.A humeral stem, head, and glenoid were explanted.The catalogue and lot number for the humeral stem and glenoid were not documented because the information was illegible.Activity level - yes.Doi: 2010.Dor: (b)(6) 2024.Affected side: left shoulder.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine, if a lot related issue was possible.A worldwide complaint database search was performed.A worldwide complaint database search found, no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude, that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected and distributed to approved specifications.Additional complaint information, monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|
|