MAUDE Adverse Event Report: MEGADYNE MEDICAL PRODUCTS, INC. SMOKE EVACUATOR; MEGADYNE¿ SMOKE EVACUATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEGADYNE MEDICAL PRODUCTS, INC. SMOKE EVACUATOR; MEGADYNE¿ SMOKE EVACUATOR Back to Search Results
Catalog Number MESE1
Device Problems Difficult or Delayed Activation (2577); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
It was reported that during the comprehensive inspection we connected it to mega power1000 and confirmed that it did not stop after it was connected.We have confirmed that the pump is making an abnormal noise.There was no adverse consequence to the patient.The customer does not want to provide additional information.
 
Manufacturer Narrative
(b)(4).Date sent: 1/11/2024 b3: only event year known: 2023 per service manual operational and diagnostic analysis confirmed the continuous activation and the noise issue.The scroll pump was replaced to address the noise issue and the pcb contact board assembly was replaced to address the continuous activation.The repair and testing of the unit was completed per the service manual, bringing the unit back to full functionality.The unit passed all functional tests and is fully operational.The repair and testing of the unit was completed.The device history records were reviewed and certed by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMOKE EVACUATOR
Type of Device
MEGADYNE¿ SMOKE EVACUATOR
Manufacturer (Section D)
MEGADYNE MEDICAL PRODUCTS, INC.
4545 creek road
cincinnati OH 45242
Manufacturer (Section G)
MEGADYNE MEDICAL PRODUCTS, INC.
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
PR  
3035526892
MDR Report Key18492131
MDR Text Key333207782
Report Number1721194-2024-00004
Device Sequence Number1
Product Code FYD
UDI-Device Identifier10614559104545
UDI-Public10614559104545
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMESE1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2023
Initial Date Manufacturer Received 12/26/2023
Initial Date FDA Received01/11/2024
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-