Model Number DBEC-125
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 09/21/2022
Event Type  Injury  
Manufacturer Narrative
H6 investigation conclusion code 22: the diamondback 360® coronary orbital atherectomy system instructions for use manual states that thrombus is a potential adverse event that may occur and/or require intervention with use of the system.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: (b)(4).
 
Event Description
Optical coherence tomography (oct) imaging detected thrombus in the mid left anterior descending artery (lad) after stenting and atherectomy using a diamondback 360 coronary orbital atherectomy device (oad).The clinical event committee reviewed the images and concluded that the use of the oad may have contributed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM
Type of Device
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
aaron stevens
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key18493535
MDR Text Key332671431
Report Number3004742232-2024-00038
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberDBEC-125
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/04/2023
Initial Date FDA Received01/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age56 YR
Patient SexMale
Patient Weight90 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-