• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2R
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, the videoscope only displayed blurry lines, no image, when the scope was plugged into the camera box.The issue was found at the beginning of a diagnostic cysto ureteroscopy.The procedure was delayed 5 minutes, and was completed using a similar device.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned to olympus for evaluation; however, the physical device evaluation has not been completed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's device evaluation and investigation.Based on the results of the device evaluation, the reported event was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.The event can be prevented by following the instructions for use which state: inspection of the endoscopic image.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18494260
MDR Text Key332736939
Report Number9610595-2024-00722
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343612
UDI-Public04953170343612
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/19/2023
Initial Date FDA Received01/11/2024
Supplement Dates Manufacturer Received03/28/2024
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OLYMPUS CYSTO SCOPE.
-
-