Model Number BREV100
Device Problem Failure to Obtain Sample (2533)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2023
Event Type  Injury  
Event Description
It was reported that during a brevera procedure on (b)(6) 2024 during the procedure the system lost suction and the samples could not be obtained.Site was unable to retrieve the samples even after cutting the tubing.The patient will return for an additional procedure to take additional samples.No other information available.A field engineer examined the equipment and the system was performing as expected, the vacuum and aspiration were within normal limits.Customer tested the equipment and was able to perform adequate testing and sampling.No other information available.
 
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.A field engineer examined the equipment and the system was performing as expected, the vacuum and aspiration were within normal limits.Customer tested the equipment and was able to perform adequate testing and sampling.No other information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BREVERA BREAST BIOPSY SYSTEM
Type of Device
INSTRUMENT, BIOPSY
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
562 parkway
san jose 20102
CR   20102
MDR Report Key18499290
MDR Text Key332722038
Report Number1222780-2024-00014
Device Sequence Number1
Product Code KNW
UDI-Device Identifier15420045507517
UDI-Public(01)15420045507517(11)210629
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberBREV100
Device Catalogue NumberBREV100
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/26/2023
Initial Date FDA Received01/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2021
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-