BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 87035 |
Device Problem
Data Problem (3196)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Intellamap orion catheter was evaluated by boston scientific.Visual inspection noted that the magnetic sensor is broken.The steering knob and the tension control knob functioned properly on both lock and unlock positions.Deployment slider functioned properly, and the basket shape was formed.No abnormal resistance was felt when executing the steering mechanism and deployment system.Analysis of returned product revealed that the magnetic sensor is broken causing that this component was found electrical open, consequently confirming the reported event.The most probable conclusion code is "cause traced to component failure" since the failure found indicates the possibility that the magnetic sensor is in malfunction or damaged.
|
|
Event Description
|
It was reported that during a cavotricuspid isthmus ablation procedure a intellamap orion catheter was selected for use.When the catheter was introduced into the body signal were visible, however, the orion could not be visualized and the console displayed the error message "error 114-2: the orion magnetic sensor is either disconnected or broken".The signal station was restarted and the umbilical cable replaced, but the issue persisted.The catheter was replaced and the procedure was completed successfully.No patient complications were reported.The device has been received at boston scientific post market laboratory.Analysis of the returned device found that the magnetic sensor is physically broken.
|
|
Search Alerts/Recalls
|
|
|