• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 87035
Device Problem Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
Intellamap orion catheter was evaluated by boston scientific.Visual inspection noted that the magnetic sensor is broken.The steering knob and the tension control knob functioned properly on both lock and unlock positions.Deployment slider functioned properly, and the basket shape was formed.No abnormal resistance was felt when executing the steering mechanism and deployment system.Analysis of returned product revealed that the magnetic sensor is broken causing that this component was found electrical open, consequently confirming the reported event.The most probable conclusion code is "cause traced to component failure" since the failure found indicates the possibility that the magnetic sensor is in malfunction or damaged.
 
Event Description
It was reported that during a cavotricuspid isthmus ablation procedure a intellamap orion catheter was selected for use.When the catheter was introduced into the body signal were visible, however, the orion could not be visualized and the console displayed the error message "error 114-2: the orion magnetic sensor is either disconnected or broken".The signal station was restarted and the umbilical cable replaced, but the issue persisted.The catheter was replaced and the procedure was completed successfully.No patient complications were reported.The device has been received at boston scientific post market laboratory.Analysis of the returned device found that the magnetic sensor is physically broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key18499482
MDR Text Key333202198
Report Number2124215-2023-75484
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number87035
Device Catalogue Number87035
Device Lot Number0031800724
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received01/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/12/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-