Model Number URF-V3 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device has been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.The customer provided the cleaning sterilization and disinfection (cds) processes performed at the user facility.The customer did confirm that there were no deviations or deficiencies concerning reprocessing of the scope.Additionally, the customer confirmed that there were no suspected patient infections due to the facility findings.The device passed the leak test.The detergent used was anios.The device was rinsed before manual disinfection.All channels were flushed with and immersed into the disinfectant.The concentration and expiration date of the disinfectant were controlled.The disinfectant used was anioxyde 1000.The water filter was replaced periodically in accordance with the instructions for use.The device was stored in simple cabinet.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
|
Event Description
|
The customer reported to olympus, during reprocessing, the uretero-reno videoscope tested positive for 6 colony forming units (cfus) of staphylococcus epidermidis, micrococcus luteus/lylae, and moraxella sp.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer completed cleaning disinfection and sterilization (cds) checklist, results of third-party testing, the device evaluation and the lms final investigation.The lm reviewed the customer provided the cds processes where no information to judge deviations from the instructions for use (ifu) was identified.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: jan 16, 2024.Sampling from: all channels.Cfu: <1 cfu/endoscope.Bacterial identification: n/a.The device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|