Catalog Number 491346
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/16/2023
Event Type  malfunction  
Event Description
It was reported when using the bd totalys slideprep, a semicircular contaminated trajectory formed on one (1) specimen.There was no report of patient impact.
 
Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Manufacturer Narrative
Investigation summary: a complaint of "a semicircular trajectory is formed on the specimen" was received against totals slideprep ce material number: 491346, serial number: (b)(6).Field service engineer (fse) investigated and concluded that there is no problem with the instrument and the instrument is working smoothly.This complaint is not a confirmed failure of the instrument.Device history record review for the instrument ppla300234 is not required for this complaint as this complaint does not allege an early life failure or failure at installation.Service history review was performed for the instrument ppla300234, and no additional work orders were observed for the complaint failure mode reported.Samples were not received by quality for investigation and thus, returned material investigation could not occur.No new trends, risks, or hazards were identified as a result of the complaint.Bd quality will continue to closely monitor for trends associated with this complaint.
 
Event Description
It was reported when using the bd totally slideprep, a semicircular contaminated trajectory formed on one (1) specimen.There was no report of patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TOTALYS SLIDEPREP
Type of Device
PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED
Manufacturer (Section D)
TRIPATH IMAGING, INC
1022 corporate park drive
mebane NC 27302
Manufacturer (Section G)
TRIPATH IMAGING, INC
1022 corporate park drive
mebane NC 27302
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18503034
MDR Text Key333206598
Report Number3008007472-2024-00001
Device Sequence Number1
Product Code MKQ
UDI-Device Identifier00382904913465
UDI-Public(01)00382904913465
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P970018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number491346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/16/2023
Initial Date FDA Received01/12/2024
Supplement Dates Manufacturer Received04/22/2024
Supplement Dates FDA Received05/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-