MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE VR; Pulse generator, permanent, implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE VR; Pulse generator, permanent, implantable Back to Search Results
Model Number CD1411-36C
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 01/05/2024
Event Type  malfunction  
Manufacturer Narrative
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found a malfunction of premature battery depletion.Interrogation of the device revealed that device was at end of life (eol) when received.A longevity calculation was performed and found the battery depletion was premature based on the device usage.Electrical testing revealed high current in the hybrid.The cause of the high current was due to a hybrid anomaly.The root cause of the hybrid anomaly could not be conclusively determined.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE VR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18503902
MDR Text Key332788833
Report Number2017865-2024-00681
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507707
UDI-Public05414734507707
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2015
Device Model NumberCD1411-36C
Device Lot Number4304142
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/05/2024
Initial Date FDA Received01/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-