Model Number WB91051W |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/25/2023 |
Event Type
malfunction
|
Event Description
|
A user facility returned the olympus generator, to the olympus service center for annual inspection.Upon inspection and testing of the returned device, it was observed that the generator board was damaged.This report is being submitted for the reportable malfunction found during evaluation.There was no patient involvement.
|
|
Manufacturer Narrative
|
Calibration and electrical safety testing will be carried out by a qualified third-party company.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the reportable malfunction (defective generator board) occurred due to a destroyed transformer.However, the definitive root cause was unable to be determined.Olympus will continue to monitor field performance for this device.
|
|