It was reported that the patient presented for a follow-up in clinic.It was noted that the right ventricular (rv) lead had high capture thresholds, device sensing issues in the form of r-wave amplitude variation, and varying low voltage impedance.A chest x-ray confirmed the rv lead was dislodged.The patient was asymptomatic.Upon trying to reposition the rv lead, the physician encountered difficulties when trying to retract the helix, but it eventually did retract.The physician then encountered further difficulties when the stylet could not be advanced in rv lead.The rv lead was explanted and replaced.The patient was stable throughout the procedure.
|
The reported events were lead dislodgement, unacceptable threshold, r-wave amp variation, varying low voltage impedance, helix mechanism issue, and failure to advance stylet.As received, a complete lead was returned in one piece with the helix found retracted and clogged with blood/ tissue.The reported event of helix mechanism issue and stylet could not be fully inserted/removed were confirmed.The cause of the helix and stylet issue was traced to the helix being clogged with blood/tissue and the inner coil being clogged with blood respectively.The reported events of unacceptable threshold, r-wave amp variation, varying low voltage impedance were not confirmed.Analysis was normal, no anomalies were found.
|