• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. CER BIOLOXD OPTION HD 36MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. CER BIOLOXD OPTION HD 36MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER Back to Search Results
Model Number N/A
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that they operator observed a scratch on the ceramic head.This event is related to a malfunction that could potentially lead to a serious injury.However, no patient harm or further outcome was reported.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified that the part was not returned in original packaging and was returned with the trial.Ceramic head shows slight wear near the taper area and there is a black mark on the side.A review of the device manufacturing records confirmed no abnormalities or deviations.Device is used for treatment.Medical records were not provided.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CER BIOLOXD OPTION HD 36MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18510109
MDR Text Key332918931
Report Number3002806535-2024-00019
Device Sequence Number1
Product Code LZO
UDI-Device Identifier00887868271397
UDI-Public(01)00887868271397(17)330726(10)3166185
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192683
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number650-1057
Device Lot Number3166185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received01/15/2024
Supplement Dates Manufacturer Received02/02/2024
Supplement Dates FDA Received02/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-